CLIA Kit for Vascular Endothelial Growth Factor Receptor 1 (VEGFR1)

Cat# USCN-SCB818Hu-96

Size : 96T

Brand : USCN

Request more information

Contact local distributor :


Phone : +1 850 650 7790
CLIA Kit for Vascular Endothelial Growth Factor Receptor 1 (VEGFR1)

CLIA Kit for Vascular Endothelial Growth Factor Receptor 1 (VEGFR1)

Flt1; Fms-Related Tyrosine Kinase 1; Vascular Permeability Factor Receptor; Fms-like tyrosine kinase 1; Tyrosine-protein kinase receptor FLT

  • Product No.SCB818Hu
  • Organism SpeciesHomo sapiens (Human) Same name, Different species.
  • Test MethodDouble-antibody Sandwich
  • Assay Length2h, 40min
  • Detection Range27.4-20,000pg/mL
  • SensitivityThe minimum detectable dose of this kit is typically less than 10.5pg/mL.
  • Sample TypeSerum, plasma, tissue homogenates, saliva, cell lysates, follicular fluid and other biological fluids
  • Download Instruction Manual
  • UOM 48T96T 96T*5 96T*10 96T*100
  • CLIA Kit for Vascular Endothelial Growth Factor Receptor 1 (VEGFR1) Packages (Simulation)
  • CLIA Kit for Vascular Endothelial Growth Factor Receptor 1 (VEGFR1) Packages (Simulation)
  • CLIA Kit for Vascular Endothelial Growth Factor Receptor 1 (VEGFR1) Results demonstration
  • SCB818Hu.jpg Typical Standard Curve
  • Certificate ISO9001: 2008, ISO13485: 2003 Registered

Specificity of the CLIA Kit for Vascular Endothelial Growth Factor Receptor 1 (VEGFR1)

This assay has high sensitivity and excellent specificity for detection of Vascular Endothelial Growth Factor Receptor 1 (VEGFR1).
No significant cross-reactivity or interference between Vascular Endothelial Growth Factor Receptor 1 (VEGFR1) and analogues was observed.

Recovery of the CLIA Kit for Vascular Endothelial Growth Factor Receptor 1 (VEGFR1)

Matrices listed below were spiked with certain level of recombinant Vascular Endothelial Growth Factor Receptor 1 (VEGFR1) and the recovery rates were calculated by comparing the measured value to the expected amount of Vascular Endothelial Growth Factor Receptor 1 (VEGFR1) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 79-104 87
EDTA plasma(n=5) 93-104 101
heparin plasma(n=5) 85-98 90

Precision of the CLIA Kit for Vascular Endothelial Growth Factor Receptor 1 (VEGFR1)

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Vascular Endothelial Growth Factor Receptor 1 (VEGFR1) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Vascular Endothelial Growth Factor Receptor 1 (VEGFR1) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity of the CLIA Kit for Vascular Endothelial Growth Factor Receptor 1 (VEGFR1)

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Vascular Endothelial Growth Factor Receptor 1 (VEGFR1) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 79-101% 81-93% 95-103% 82-101%
EDTA plasma(n=5) 98-105% 99-105% 81-93% 94-105%
heparin plasma(n=5) 84-101% 78-99% 82-98% 94-102%

Stability of the CLIA Kit for Vascular Endothelial Growth Factor Receptor 1 (VEGFR1)

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Assay procedure summary of the CLIA Kit for Vascular Endothelial Growth Factor Receptor 1 (VEGFR1)

1. Prepare all reagents, samples and standards;
2. Add 100µL standard or sample to each well. Incubate 1 hours at 37°C;
3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 100µL Substrate Solution. Incubate 10 minutes at 37°C;
8. Read RLU value immediately.

Test principle of the CLIA Kit for Vascular Endothelial Growth Factor Receptor 1 (VEGFR1)

The microplate provided in this kit has been pre-coated with an antibody specific to Vascular Endothelial Growth Factor Receptor 1 (VEGFR1). Standards or samples are then added to the appropriate microplate wells with a biotin-conjugated antibody specific to Vascular Endothelial Growth Factor Receptor 1 (VEGFR1). Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. Then the mixture of substrate A and B is added to generate glow light emission kinetics. Upon plate development, the intensity of the emitted light is proportional to the Vascular Endothelial Growth Factor Receptor 1 (VEGFR1) level in the sample or standard.;